7 Reprocessing

7.4 Manual maintenance

B
A
A
B
16
Manual maintenance
7.4
7.4.1

Maintenance of mechanical parts

Intervals
● Prior to each sterilization
● After every thermal disinfection without integrated maintenance
Required accessories
A
B
C
NOTE!  Use only Dentsply Sirona T1 Spray.
C
Procedure
ü The spray adapter is disinfected.
ü The O-ring on the spray adapter is intact.
1. Fit the spray adapter onto the nozzle of the spray can.
2. Insert the instrument until it snaps into place and hold it.
3. Spray the instrument for 1 - 2 seconds.
IMPORTANT: Hold the spray can upright.
4. Wipe any spray that comes out with a disinfection cloth.
5. Repeat the process until the spray that comes out of the instrument
is clear.
7.4.2

Care of the push button chuck

Use Dentsply Sirona T1 Spray on the push-button chuck to remove
deposits and ensure proper functioning of the clamping system.
Interval
● At least once a week
Required accessories
A
B
NOTE!  Use only Dentsply Sirona T1 Spray.
Procedure
ü The instrument is clean and disinfected.
1. Press the contra-angle handpiece head with the chuck firmly
against the spray can nozzle.
2. Spray the chuck for 1 - 2 seconds.
IMPORTANT: Hold the spray can upright.
3. Wipe any spray that comes out with a disinfection cloth.
Spray adapter
O-ring for spray adapter
T1 Spray
Black nozzle insert
Dentsply Sirona T1 Spray
Dentsply Sirona
Instructions for use VDW 6:1
61 34 378 D3451
D3451.201.04.07.02    07.2019