Hach ORBISPHERE 51 Series User Manual page 41

With k1200 and tc sensors
Hide thumbs Also See for ORBISPHERE 51 Series:
Table of Contents

Advertisement

Installation
The flow chamber should be mounted in such a way that the sample outlet port is located at the
lowest point to allow condensation to escape with the outgoing gas. Attach the flow chamber to
a vertical support with the screws supplied. The pressure sensor must be on top.
Note: A user manufactured spacer (~15 mm thick) may be used between the flow chamber and support for
improved access for sensor removal.
1. Top view
2. Sensor
3. Optional second sensor
4. Sample inlet
40
Figure 24 32002.xxx multi parameter flow chamber
Shown here with: gas sensor (right), pressure sensor (center) and optional second sensor (left)
5. Front view
6. Pressure sensor housing
7. Sample outlet