1. Press "SETUP"
Controller will prompt for a
password. A flashing cursor on
the entry screen will make it clear
which digit is being changed. Use
the "
" keys to scroll to the
appropriate number. Press
"ENTER" to advance the cursor to the next position.
2. Press "ENTER"
The display will read
"Temperature Menus".
3. Press "SETUP"
Until the display reads "Liquid
Level Menus".
4. Press "ENTER"
The display will read "High Level
Alarm". Use the "
" keys
to adjust the High Level Alarm
value and press "ENTER" to
save the change.
5. Press "SETUP"
The display will read "High Level
Setpoint". Use the "
to adjust the High Level Setpoint
value and press "ENTER" to save
the change.
6. Press "SETUP"
The display will read "Low Level
Setpoint". Use the "
to adjust the Low Level Setpoint
value and press "ENTER" to save
the change.
7. Press "SETUP"
The display will read "Low Level
Alarm". Use the "
" keys to
adjust the Low Level Alarm value
and press "ENTER" to save the
change.
Enable / Disable Auto Fill Control
Auto Fill Control can be enabled or disabled in the Advanced
Level Menus.
Note: Security Level 3 or higher is required to enable / disable
Auto Fill.
Thermo Fisher Scientific CryoExtra CE8100 Series
Higher User Level
Required use
ENTER Password 0000
Press ENTER for
Temperature Menus or
press SETUP for next
menu
Press ENTER for
Liquid Level Menus or
press SETUP for next
menu
High Level Alarm
8.0 in Use
to
adjust
Press ENTER to save
High Level Setpoint
7.0 in Use
" keys
adjust
Press ENTER to save
Low Level Setpoint
" keys
5.0 in
Use
to adjust
Press ENTER to save
Low Level Alarm 4.0 in
Use
to adjust
Press ENTER to save
1. Press "SETUP"
Controller will prompt for a
password. A flashing cursor on the
to
entry screen will make it clear
which digit is being changed. Use
the "
" keys to scroll to the appropriate number. Press
"ENTER" to advance the cursor to the next position.
2. Press "ENTER"
The display will read "Temperature
Menus".
3. Press "SETUP"
Until the display reads "Liquid
Level Menus".
4. Press "ENTER"
The display will read "High Level

Alarm".

5. Press "SETUP"
Until the display reads "Advanced
Level Menus".
6. Press "ENTER"
The display will read Auto Fill
to
Control". Use the "
adjust and press "ENTER" to save
the change.
Level Offset
Since the annular line that carries the pressure level signal to
the Controller is not located at the very bottom of the LN
column, a Level Offset is needed. This value is determined
and set at the factory and should not be adjusted to fix an
inaccurate level reading unless setting up a freezer (see the
"First Fill Startup procedure") or if instructed to do so by
an authorized CryoExtra Distributor or Technical Service. For
liquid level calibration see section "Liquid Nitrogen Level
Calibration".
Note: Security Level 3 or higher is required to adjust the Level
Offset.
1. Press "SETUP"
Controller will prompt for a
password. A flashing cursor on the
entry screen will make it clear
which digit is being changed. Use
the "
" keys to scroll to the
Higher User Level
Required use
ENTER Password 0000
Press ENTER for
Temperature Menus or
press SETUP for next
menu
Press ENTER for
Liquid Level Menus or
press SETUP for next
menu
High Level Alarm
8.0 in
Use
to adjust
Press ENTER to save
Press ENTER for
Advanced Level Menus
or press SETUP for next
menu
Auto Fill Control
" keys to
ENABLED
Use
to adjust
Press ENTER to save
Higher User Level
Required
use
to ENTER
Password 0000
Operation |
to
2
27