Internal Calibration Performance Verification
This performance verification is carried out to verify the accuracy of the U8480
Series internal calibration.
This performance verification requires the following equipment:
– power meter (N1914A)
– computer, with the BenchVue Power Meter/Power Sensor Control & Analysis
Procedure
1 Turn on the power meter, and connect the DUT (U8480 Series) to the
2 Launch the Keysight IO Libraries Suite on the computer, and connect the DUT
3 Perform internal calibration for the DUT by sending "CAL:TYPE INT" followed
4 Change the power measurement unit of the DUT to watt by sending
5 Turn on the 1 mW reference calibration source of the power meter, and then
6 Set the DUT to the single trigger mode by sending "INIT:CONT OFF".
Keysight U8480 Series Service Guide
App and the Keysight IO Libraries Suite installed
computer.
to the 1 mW reference calibration connector of the power meter. The
equipment setup is as shown in the figure.
by "CAL:ALL".
"UNIT:POW W".
wait for 3 hours.
Performance Verification and Adjustments
Power meter
DUT
2
29