Performance Verification and Adjustments
7 Read the power level of the DUT by sending "READ?", and then record the
8 Compare the reading to the specification in the datasheet. If the verification
30
reading.
fails, repeat steps 3 to 7. If the verification fails for three times consecutively,
refer to
"Adjustments"
on page 31.