Keysight Technologies M9216A Startup Manual page 19

Pxi high voltage data acquisition
Table of Contents

Advertisement

M9216A Startup Guide
Keysight M9216A PXI High Voltage Data Acquisition
Figure 3
Installing the module to the chassis
5 Hold the module by the injector/ejector handle, and slide
it into an available PXI (or hybrid) slot, as shown in
Figure
3.
• Install the module into the PXI slot of the chassis by
placing the module card edges into the front module
guides (top and bottom).
• Slide the module to the rear of the chassis and assure
that the injector/ejector handle is pushed down in the
unlatched (downward) position.
• Slide the module completely into the chassis.
• When you begin to feel resistance, push up on the
injector/ejector handle to fully inject the module into
the chassis.
6 If you are using a remote controller, install the System
Interface Card in the chassis.
7 Latch the module by pulling up on the injector/ejector
handle and secure the front panel to the chassis using the
module front- panel mounting screws.
8 Tighten the screws on the module (or remote controller)
front panel. Performance may suffer if the screws are not
tightened properly.
9 Verify that the PXI chassis fans are operable and free of
dust and other contaminants that may restrict airflow.
11

Advertisement

Table of Contents
loading

Table of Contents