BW RigRat User's Guide
During startup, sensors and functions are checked. Also, if sensors are overdue for a bump test or
calibration, either or both of these screens will be displayed to alert you:
Note: If you do not click the button to acknowledge being informed of overdue tests, the instrument will
shut itself off in 300 seconds (5 minutes).

Turning The RigRat Off

At the main measurement screen, press and hold the button. A 3-second countdown to shutoff begins.
After that, another 3-second countdown to enter the menu begins. You must release the button during
the second 3-second countdown to shut off the instrument. The screen displays "OFF" for 3 seconds,
and then the instrument is off. If you release the button during the first 3-second countdown, the
instrument continues normal operation.
Note: if you continue to hold the button during the second 3-second countdown, it enters the main
menu.
Note: You cannot turn off the RigRat if it is in gas alarm.
46