Teledyne 5800 Removal And Replacement page 2

115v refrigeration module

Advertisement

Instruction Sheet 60-4702-048 Revision H, November 8, 2019
4. Remove the refrigeration module mounting
screws (Figure 3).
5. Carefully pull the module out and rotate
clockwise to expose the wiring connectors. The
bushing holding the AC line cord in place should
slide along the cord (see inset, Figure 3). During
reinstallation, the slack created in this step must
be removed.
CAUTION
Keep the module as close to the refrigerator body as
possible to avoid pulling the wiring taut and damaging
the connectors.
Pull module out and
rotate clockwise
Line cord bushing
Figure 3: Remove refrigeration module mounting
screws
6. Remove the terminal cover of the compressor
(Figure 4). For serial number 219K02629 or
newer, remove the screw and lock washer that
hold the green/yellow ground wire in place on
the compressor.
Figure 4: Access compressor overload A screwdriver
may be used to release the latch.
7. Disconnect the four (4) compressor wires from
the power supply cable (Figure 5).
a. While holding the overload relay in place,
disconnect the brown wire.
b. Disconnect the blue wire from the capacitor
and relay assembly from the gray wire on the
power supply cable.
c. Disconnect the black heater wire from the
black wire on the power supply cable.
11
d. With the
/
" nut driver or socket wrench,
32
remove the nut holding the green/yellow
ground wire on the module chassis.
Blue
Gray
Black
Heater
Wire
Figure 5: Disconnect compressor wires
8. The evaporator temperature sensor is mounted
on the sixth coil up on the evaporator (Figure 6)
and has a black rubber molded housing. Cut the
ties holding this sensor in place and move it out
of the frame.
Figure 6: Remove evaporator temperature sensor
Brown
Green/
yellow