15B & 17B
Calibration Manual

General Specifications

Maximum Voltage between any
Terminal and Earth Ground:
W Fuse Protection for mA µA inputs:
W Fuse Protection for A input:
Display:
Temperature:
Operating Altitude:
Temperature Coefficient:
Electromagnetic Compatibility:
Rated Transient Overvoltage:
Relative Humidity:
Relative Humidity, 40 MΩ Range:
Battery Type:
Battery Life:
Size (H x W x L):
Weight:
Certifications:
Safety Compliance:
Overvoltage installation categories per IEC 61010-1, 2000: The Meter is designed to protect against
transients in these catagories:
CAT I: From high-voltage low-energy sources, e.g., electronic circuits or a copy machine.
CAT II: From equipment supplied from the fixed installation, e.g., Televisions, PCs, portable tools and
household appliances.
CAT III: From equipment in fixed equipment installations, e.g., installation panels, feeders and short
branch circuits, and lighting systems in large buildings.
4
1000 V
500 mA, 1000 V FAST, Min. Interrupt Rating 20,000 A
10 A, 1000 V FAST, Min. Interrupt Rating 20,000 A
Digital: 4000 count updates 3/sec
Operating: 0 °C to 40 °C,
indefinitely (to -40 °C for 100 hours)
0 to 2000 meters
0.1 X (specified accuracy)/ °C (<18 °C or >28 °C)
Complies with FCC Part 15, Class B, IEC 61326, 3 V/m,
performance criterion B
4 kV (1.2 x 50 µs) Peak for measurement Categories I, II,
and III.
Noncondensing < 10 °C
90% from 10 °C to 30 °C; 75% from 30 °C to 40 °C
80% from 10 °C to 30 °C; 70% from 30 °C to 40 °C
2 X AA, NEDA 15A, IEC LR6
Alkaline: 500 Hours
180 mm x 89 mm x 51.5 mm (with holster)
425 grams
CMC, P
IEC 61010-1, 2000 CAT I 1000 V, CAT II 600 V, and CAT III
300 V overvoltage standards
Storage: -30 °C to 60 °C