Performance Test - Fluke 15B Calibration Manual

Hide thumbs Also See for 15B:
Table of Contents

Advertisement

Test
Step
Function
v
1
Volts AC
2
3
4
5
6
7
8
9
10
11
12
13
Set 5520 to Standby.
Table 4. 17B Performance Test
UUT Switch
UUT Push
Setting
Button
N/A
R
R
R
I
-
-
-
-
I
5520A Output
UUT Display limits
3.5 V 50 Hz
3.462 to 3.538 V ac
35 V 500 Hz
34.62 to 35.38 V ac
350 V 50 Hz
346.2 to 353.8 V ac
1000 V 50 Hz
Auto Range icon
OFF
OL.
987 to 1013 V ac
0.35 V 50 Hz
339.2 to 360.8 mV ac
1 V 50Hz
49.92 to 50.08 Hz
1 V 500 Hz
499.2 to 500.8 Hz
1 V 5 kHz
4.992 to 5.008 Hz
2 V 50 kHz
49.92 to 50.08 Hz
3 V 100kHz
99.6 to 100.4 Hz
3 V 500 Hz
49.0 to 51.0 %
Digital Multimeter
Performance Tests
11

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

17b

Table of Contents