Test
Step
Function
For step 35, wait 15 seconds for the reading to settle.
c
35
Capacitance
A
36
Amps DC
37
A
38
Amps AC
39
Set 5520A to Standby.
m
40
mA dc
41
42
mA ac
43
Set 5520A to Standby.
N
µA dc
44
45
µA ac
46
47
Set the 5520A to Standby. Connect 5520A temperature output to UUT V Ω terminals using 80AK and K
type Thermocouple wire with male mini connectors at each end. Refer to Figure 2.
Wait 3 to 5 minutes before performing steps 48 and 49 to allow test lead thermals to dissipate.
d
48
Temperature
49
Performance testing of the 17B is now complete.
Table 4. 17B Performance Test (cont.)
UUT Switch
UUT Push
Setting
Button
-
N/A
C
-
C
-
-
C
-
-
-
5520A Output
UUT Display limits
95 µF
t icon ON
89.7 to 100.3 µF
3.8 A dc
3.713 to 3.887 A dc
-10 A dc
-9.82 to -10.18 A dc
10 A 50 Hz
9.82 to 10.18 A ac
3.8 A 50 Hz
3.713 to 3.887 A ac
38 mA
37.40 to 38.60 mA dc
0.38 A dc
374.0 to 386.0 mA dc
0.38 A 50 Hz
374.0 to 386.0 mA ac
38 mA 200 Hz
37.40 to 38.60 mA ac
3740 to 3860 µA dc
3.8 mA
374.0 to 386.0 µA dc
0.38 mA
374.0 to 386.0 µA ac
0.38 mA 40 Hz
3740 to 3860 µA ac
3.8 mA 200 Hz
380 °C
371.4 to 388.6 °C
-55.0 °C
-48.0 to -62.0 °C
Digital Multimeter
Performance Tests
13