15B & 17B
Calibration Manual
Step
Function
14
Volts DC
15
16
17
18
19
20
mV dc
21
Ohms
22
23
24
25
26
27
Diode
28
Continuity
29
For step 30, disconnect test leads from the calibrator first, then push
This 'rels out' test lead and meter capacitances.
30
Capacitance
31
32
33
For step 34, wait 15 seconds for reading to settle.
34
Capacitance
12
Table 4. 17B Performance Test (cont.)
Test
UUT Switch
Setting
V
M
F
c
c
UUT Push
5520A Output
Button
N/A
3.5 V dc
35 V dc
350 V dc
1000 V dc
r
STANDBY
N/A
0.35 V dc
350 Ω
3.5 kΩ
35 kΩ
350 kΩ
3.5 MΩ
35 MΩ
C
0.7 V dc
(once)
40 Ω
C
(press again)
800 Ω
-
r
r
STANDBY
-
48 nF
0.48 µF
-
4.8 µF
-
48 µF
-
UUT Display limits
3.480 to 3.520 V dc
34.80 to 35.20 V dc
348.0 to 352.0 V dc
992 to 1008 V dc
t icon ON
-992 to -1008 V dc
345.5 to 354.5 mV dc
347.9 to 352.1 Ω
3.480 to 3.520 kΩ
34.80 to 35.20 kΩ
348.0 to 352.0 kΩ
3.480 to 3.520 MΩ
34.45 to 35.55 MΩ
D icon ON
0.630 to 0.770 V
E Icon ON
Beeper On
Beeper OFF
when the reading settles.
t icon ON
0.00 to 0.02 nF
t icon ON
46.99 to 49.01 nF
t icon ON
469.9 to 490.1 µF
t icon ON
4.555 to 5.045 nF
t icon ON
45.55 to 50.45 µF