Test
Step
Function
v
1
Volts AC
2
3
4
5
6
7
Set 5520A to Standby
8
Volts DC
9
10
11
Set 5520A to Standby
M
12
mV dc
F
13
Ohms
14
15
16
17
18
F
19
Diode
F
20
Continuity
21
-
-
Table 3. 15B Performance Test
Switch
Push Button
Position
N/A
R
R
R
N/A
N/A
N/A
C
(once)
C
(press again)
-
5520A
UUT Reading
Output
3.5 V 50 Hz
3.462 to 3.538 V ac
35 V 500 Hz
34.62 to 35.38 V ac
350 V 50 Hz
346.2 to 353.8 V ac
1000 V 50 Hz
Auto Range
icon OFF
OL.
987 to 1013 V ac
0.35 V 60 Hz
339.2 to 360.8 mV ac
3.5 V dc
3.480 to 3.520 V dc
35 V dc
34.80 to 35.20 V dc
350 V dc
348.0 to 352.0 V dc
-1000 V dc
-992 to -1008 V dc
0.35 V dc
345.5 to 354.5 mV dc
350 Ω
347.9 to 352.1 Ω
3.5 kΩ
3.480 to 3.520 kΩ
35 kΩ
34.80 to 35.20 kΩ
0.35 MΩ
348.0 to 352.0 MΩ
3.5 MΩ
3.480 to 3.520 MΩ
35 MΩ
34.45 to 35.55 MΩ
D icon ON
0.7 V dc
0.630 to 0.770 V dc
40 Ω
E Icon ON
Beeper ON
800 Ω
Beeper OFF
Digital Multimeter
Performance Tests
Limits
9