Electron Capture Detector (ECD)
Replacement procedures
Replacing the detector interface card
The ECD detector interface board plugs into the main board as shown below.
1.
Unlock and unclip the ribbon cable from the interface board.
2.
Remove the screw at the top of the card where it mounts to the main
board and pull the card out.
Figure 340-13
16 of 24
Detector interface board installed in back detector position
Detectors
Agilent 6890 Gas Chromatograph Service Manual
Detector card screw
Ribbon cable
ECD detector board