Removing Both The Manifold And Plumbing Assemblies - Agilent Technologies 6890 Series Service Manual

Gas chromatograph
Hide thumbs Also See for 6890 Series:
Table of Contents

Advertisement

240
Programmed Temperature Vaporization Inlet (PTV)
Replacement procedures
Filter and O-rings
(2) (not shown)
PTV front trap
assembly
Filter and O-rings
(2) (not shown)
PTV front trap
assembly
Figure 240-15

Removing both the manifold and plumbing assemblies

1.
Loosen the three captive Torx screws which secure the inlet to the oven,
and lift the PTV inlet out of the oven.
2.
Disconnect the carrier gas supply line, split vent line, and septum purge
line (for the septum head) from the inlet. Straighten the tubing at the
manifold so that the manifold can be easily removed from the chassis.
22 of 46
PTV plumbing assemlies
Inlets
Agilent 6890 Gas Chromatograph Service Manual
PTV SMLS
gang weldment
Jun 2001