Test F-5: Interrupt/Status Information
Test F-5: Interrupt/Status Information
Description
This test uses the following commands to return information on command
module interrupt lines and on register status. See Chapter 4 - Status and
Interrupts in the Keysight E1406A Command Module User's Manual for interrupt
and status information.
Interrupt Information
Status Information
Set Up Equipment
Keysight E1406A Command Module Service Guide
DIAG:INT:SETn? State of interrupt line n
STAT:OPER:COND? State of Condition register
STAT:OPER:ENAB? Standard Operation Enable register mask value
STAT:OPER:EVEN? Value of bit set in Event register
STAT:QUES:ENAB? Questionable Status Register enable mask value
– Turn Mainframe Power OFF
– Connect computer to Command Module (see Figure 2-12)
– Turn Mainframe Power ON
Verification Tests
53
Need help?
Do you have a question about the VXI bus 75000 C Series and is the answer not in the manual?