Command Module Self-Tests

Command Module Self-Tests
This section shows how to perform Keysight E1406A Command Module
self-tests using the GPIB power-on test and/or RS-232 power-on test. Either test
is usually adequate to verify that a command module is operational. If a self-test
fails, see Chapter 4 - Service for further tests/information.

Test S-1: GPIB Power-On Test

Description

This test uses the SYST:ERR? command for the command modules GPIB
power-on test. A "+0, No error" return indicates the test passed.

Set up Equipment

Keysight E1406A Command Module Service Guide
Unless otherwise instructed, the Run/Load switch should be in
the Run position for all tests.
– Turn mainframe power OFF
– Connect computer to mainframe (see Figure 2-4)
– · Turn mainframe power ON
Refer to your computer's documentation for information on
connecting the keyboard and video cables and other peripherals.