Grounding Practices - Emerson Rosemount 848T Reference Manual

Wireless temperature transmitter
Hide thumbs Also See for Rosemount 848T:
Table of Contents

Advertisement

Installation
September 2019
3.4.2

Grounding practices

The transmitter operates with the housing floating or grounded. However, the extra noise
in floating systems may impact many types of readout devices. If the signal appears noisy
or erratic, grounding the transmitter at a single point may solve the problem.
The electronics enclosure should be grounded according to local and national installation
codes. This can be accomplished via the process connection, internal case grounding
terminal, or the external grounding terminal.
Each process installation has different requirements for grounding, use the options
recommended by the facility for the specific sensor type, or begin with the
recommendations below.
Ungrounded thermocouple, mV, and RTD/ohm inputs option
Procedure
1. Connect sensor wiring shield to terminal screw 5 at the terminal block. Terminal
2. Ensure the sensor wiring is electrically isolated from the transmitter housing.
Grounded thermocouple option
Procedure
1. Ground the sensor wiring shield at the sensor.
2. Ensure the sensor wiring and shield is electrically isolated from the transmitter
34
screw 5 is internally connected to the housing.
A
Shield ground point
housing and terminal screw 5.
Reference Manual
00809-0100-4848
A
Emerson.com/Rosemount

Advertisement

Table of Contents
loading

Table of Contents