Appendix D Safety

Safety and electromagnetic compatibility (EMC) standards

Laser safety

Safety and electromagnetic compatibility (EMC) standards
40
WARNING! LASER HAZARD. Under normal operating conditions, the
GeneTitan
Scanner is categorized as a Class 1 laser product. However,
removing the protective covers and (when applicable) defeating the interlocks
can result in exposure to the internal Class 2 (II) laser. Lasers can burn the
retina, causing permanent blind spots. To ensure safe laser operation:
·
Never look directly into the laser beam.
·
Do not remove safety labels, instrument protective panels, or defeat safety
interlocks.
·
The system must be installed and maintained by a Thermo Fisher Scientific
Technical Representative.
·
Remove jewelry and other items that can reflect a laser beam into your eyes
or those eyes of others.
·
Wear proper eye protection and post a laser warning sign at the entrance to
the laboratory if the laser protection is defeated for servicing.
·
DO NOT operate the laser when it cannot be cooled by its cooling fan; an
overheated laser can cause severe burns on contact.
The following table lists laser safety symbols and alerts that may be present on the
instrument.
CAUTION! Class 2 (II) visible and/or invisible laser radiation present when open
and interlocks defeated. Do not stare directly into the beam or view directly with
optical instruments.
MISE EN GARDE ! RISQUE LIÉ AU RAYONNEMENT LASER, Lecteur de code-
barres. Le lecteur de code-barres inclut dans l'instrument est un appareil laser
de classe 2. Pour éviter toute lésion oculaire, ne regardez pas directement le
faisceau et ne le dirigez pas vers les yeux d'une autre personne.
MISE EN GARDE ! Rayonnement laser visible ou invisible de classe 2 (II) présent
en position ouverte et avec les dispositifs de sécurité non enclenchés. Ne pas
regarder directement dans le faisceau ou avec des instruments optiques.
The instrument design and manufacture complies with the following standards and
requirements for safety and electromagnetic compatibility.
Alert
Alerte
GeneTitan
Scanner Use and Maintenance Guide