GeneTitan
Scanner Use and Maintenance Guide
Reset the lamp life counter
Using the GCC Standalone Scanner Control module accessed from the Launcher
window, you must alert the software that the lamp has been replaced so that the hours
of the lamp counter are reset. This menu option is only available when the system is
not active.
1. Select Tools4Reset Lamp Life Time.
2. Read the confirmation message that appears, then confirm that all conditions
have been met.
3. When ready, click Yes in the message window to reset the counter.
Chapter 3 Maintain the Instrument