Component clearances required for installation and maintenance
Bench
Requirements
IMPORTANT! Air flow around the instrument must not be restricted.
Depth
Width
Height
Must be a minimum of 76.2 cm (30 in.) height clearance.
Lambda LS xenon arc lamp specifications
GeneTitan
Scanner Use and Maintenance Guide
™
During installation and maintenance, it is necessary to access the front, back, and right
side of the product. If the back of the product faces a wall, ensure that there is
sufficient clearance on the bench to rotate the product.
IMPORTANT! For safety reasons, the power outlet for the product must be accessible
at all times.
• Level surface
• Capable of bearing the weight of the instrument system and components: 113.4 kg (250
lbs)
• Must be at a minimum depth of ~79 cm (31 in.) deep to hold the instrument, xenon arc
lamp, monitor, and keyboard.
• Allow for 5.1 cm (2 in) in the back for proper airflow.
• If the bench has wheels, it can facilitate access to the back of the instrument.
• Must be a minimum width of 142.2 cm (56 in.) for the instrument, xenon arc lamp,
computer, monitor, and keyboard.
• Allow for 30.5 cm × 30.5 cm (12 in × 12 in) of clearance to the left of the instrument for
array plate access.
Parameter
Output range
Radiant output
Lamp type
Lamp life
Dimensions (H x W x D)
Weight
Electrical
Mains voltage
Maximum power
consumption
Appendix C Instrument specifications
Component clearances required for installation and maintenance
Minimum clearance and requirements
320-700 nm (standard, ozone-free bulb)
50 watts (broadband, full beam) for 300 W bulb
300 W Xenon, pre-aligned to produce collimated output
Lamp warranted for 500 hours; expected lifetime is 500 hours
10.5 x 9.5 x 10 inches (26.7 x 24.1 x 25.4 cm)
10.5 lb (4.8 kg)
110 through 240 V 50 through 60 Hz
300 W
Value
C
31