Summary of Contents for hadewe Xantos-Nail 0872

  • Page 1 Manual Xantos-Nail art. 0872 FB 04-182 Rev. 5 Date: 21.01.2019...
  • Page 2: Table Of Contents

    Contents Safety Instructions Cleaning & Disinfection Getting to Know the Unit Before First Use Area of Application Installation Site Application Starting up and Operating the Unit Clamping the Bur/Tool Setting up the Speed Maintenance & Care Handling Instructions Troubleshooting Technical Data...
  • Page 3: Safety Instructions

    Safety Instructions Read this instruction manual in its entirety before you start operating the unit. Especially read through any information regarding the battery and recharging process to prevent damage to the accumulators. The equipment should only be used by trained operators. The installation of the working area has to correspond to the relevant regulations.
  • Page 4 Safety Instructions that you have unplugged the unit before. Check the mains cable regularly for damage, and in case of such, have it changed or repaired by an authorized service center. Prevent your hair from wrapping itself around moving parts. If applicable, wear a hair net. When working with materials, which might create dust or moisture, use a drilling unit with suction or spray technique.
  • Page 5 Safety Instructions This appliance can be used by children aged from 8 years and above and persons with reduced physical, sensory mental capabilities or lack of experience and knowledge they have been given supervision or instruction concerning use of the appliance in a safe way and understand the hazards involved.
  • Page 6 Safety Instructions • the motor blocks during the treatment. Regularly check the mechanical safety of the unit (look for damages to the housing or mains cable for example). Always follow the operating instructions. Do not carry out any repairs, modifications or maintenance work yourself.
  • Page 7: Cleaning & Disinfection

    Cleaning & Disinfection After every treatment please carry out the following steps: • Remove the rotating tool (e.g. bur, capper, etc.) first. Then handle the cleaning of the tool according to the instructions of its manufacturer. • Remove tissue remains by wiping off the unit with a mildly moistened paper cloth.
  • Page 8: Getting To Know The Unit

    Getting to Know the Unit 1. speed display 2. LED 3. increase speed 4. change direction of rotation 5. decrease speed 6. ON/OFF button...
  • Page 9: Before First Use

    Before First Use Please read through the instructions Installation Site carefully before using the unit and The installation of the workroom carefully note the safety advice. must correspond to any regulations Always keep this manual accessible according to your country laws. Set to any users of the drill.
  • Page 10: Application

    Application Starting up and Operating the Unit Put down the handpiece in the handpiece holder. Plug in the plug. Turn on the unit with button 6. With For security reasons, only button 3 and 5 you can adjust the change the tool if the speed level.
  • Page 11: Maintenance & Care

    Maintenance & Care Cleaning the Chuck Weekly cleaning (every 20 working hours) Screw off the top (pg. 10) and brush off the loose parts with the added brush. Hold the handpiece that way that the top points downward the table to avoid that more dirt can enter the inside of the handpiece.
  • Page 12: Handling Instructions

    In cases in which a rust Working with high pressure infestation was found within the will not improve the outcome handpiece,...
  • Page 13: Troubleshooting

    Troubleshooting When the motor is not running or carbons was added to your delivery. blocks the carbons of the motor Then also check second could be worn out… carbon. The following described work should only be carried out by If the handpiece becomes hot… trained and experienced staff.
  • Page 14 Troubleshooting Assembly Drawing Attend to fitting position! More manuals for repair workshops you will find at http://hadewe- online.de/guides/index.htm...
  • Page 15: Technical Data

    4000 - 25000 rpm 2014/30/EU directive EMC 2014/35/EU low voltage directive control unit: weight: 0,9 kg hadewe GmbH – Grambartstraße 10 – measurements: W114xH86xD183mm 30165 Hannover – Germany power supply: 230V ~, 50Hz output: 50 W Safety Check fuse pri.:315mA delayed-action fuse...
  • Page 16 GmbH Grambartstraße 10 30165 Hannover Germany info@hadewe.de www.hadewe.de...