Table of Contents
loading

Summary of Contents for hadewe Hyperios4 PLUS

  • Page 1 Manual Hyperios PLUS Art. 0586 FB 04-295 Rev. 1 Date: 28.01.2019...
  • Page 2: Table Of Contents

    Directory Foreword Safety Instructions Cleaning & Disinfection Getting to Know the Unit Before First Use Area of Application Operation Application Operation Steps Work Positioning Infilling the Spray Liquid Insert Bur Adjusting the Speed Memory Buttons m1, m2 Spray Drying Function/Changing the Bur Foot Switch Maintenance &...
  • Page 3: Foreword

    Foreword The Hyperios Plus foot care unit was produced according to strict quality criteria and uses the latest technology in its design. The images in this manual may vary slightly from the product you have bought as we are constantly developing our products. This user manual includes a detailed description and explanation of how to use the Hyperios Plus unit.
  • Page 4: Safety Instructions

    Safety Instructions The equipment should only be used by trained operators, who know the function of the unit well. The installation of the working area has to correspond to the relevant regulations. Set up the unit in such a way, that the louvers are not covered and the extracted air can escape well.
  • Page 5 Safety Instructions with a damped cloth. Be sure you unplug the unit before. Only when pulling the mains cable, you can ensure that the device will be completely disconnected from mains. Check the mains cable regularly for damages, and in case of such, let it change or repair from an authorized service center.
  • Page 6 Only use distilled water plus a sterilizing additive (e.g. hadewe AquaHy, art. 8555). If you use other brands, pay attention to their manuals if you can use them for this purpose.
  • Page 7 Safety Instructions Change the spray liquid everyday. Always add disinfection additive to the distilled water. Rinse the tank daily and fill it with fresh water everyday. To ensure that the burr/ tool sits firmly in the chuck, do not work with tools which: •...
  • Page 8 Safety Instructions Empty the unit if it is not in use for a longer period of time (e.g. before vacations). Clean out the tank. Set the spray function on maximum. Now start the spray function, and the water will spray for approx. 3 min. The fluid will be blown out of the complete spray system and handpiece.
  • Page 9: Cleaning & Disinfection

    Disinfection Cleaning & The device is designed to be compatible with rotating in- struments which do not penetrate the skin or mucosal or inadvertently may still cause injuries. For this application please adhere to the cleaning and disinfection instructions provided be- low.
  • Page 10: Getting To Know The Unit

    Getting to Know the Unit 10 11 12 13 14 1 spray regulator knob 2 turn OFF/ON spray 3 decrease speed 4 increase speed 5 LED lights up during anticlockwise operation 6 button to change the direction of rotation 7 turn off handpiece & spray/turn on airflow 8 LED lights up when the washer change necessary 9 button to confirm the changing the washer 10 handpiece light ON/OFF...
  • Page 11: Before First Use

    Before First Use Please read through the instructions Operation carefully before using the unit and The working area must conform to carefully note the safety advice. the conditions described. Set up the Always keep this manual accessible unit in such a way, that the air slits to any users of the drill.
  • Page 12 Before First Use If you have a foot switch, insert its plug into the port on the right of the device which displays this symbol Make sure, that the tank is filled with spray liquid. Ensure that the tool is fixed. On the back of the device you will find the main power switch.
  • Page 13: Application

    Application Operation Steps and spray will turn off. The hand- piece can now be placed in its In the following section the order in holder on the right of the unit. Alter- which you should use the device is natively, you can switch the device described.
  • Page 14: Work Positioning

    Application Work Positioning The following images are examples of how to work on various parts of the foot. Always make sure that the foot is firmly gripped and that you can work freely and without strain.
  • Page 15 Application...
  • Page 16: Infilling The Spray Liquid

    Application Infilling the Spray Liquid Insert Bur Switch off the device. Pull out the The handpiece is equipped with a spray tank on its handle. Hold the chuck which can hold all burs with a cover of the tank in hand and un- diameter of 2.35 mm and usually screw the tank.
  • Page 17: Adjusting The Speed

    4 seconds. To recall the above speed Commencement of foot switch level simply press button m1 (11) Only use the original hadewe foot again. Further speeds can be saved switch to avoid damages. Insert the with m2 (12).
  • Page 18: Maintenance & Care

    Maintenance & Care Washer Water Filter handpiece Even smallest parts of dirt could equipped with a white washer (art. block the nozzle, which sits in the 4571). This washer has to be re- top. Avoiding such blockages, there placed by a new one every 50 op- is a water filter installed in the cap eration hours, unless exchanging of the tank.
  • Page 19: Cleaning Function

    Maintenance & Care Cleaning Function mance of the handpiece. Therefore, clean the chuck monthly. Use the cleaning function if the nozzle is blocked. Turn the unit off using button 14 (Standby). Remove How to remove the chuck the water filter at first. (For its in- structions see paragraph above.) Screw off the top of the handpiece Fill the tank with distilled water.
  • Page 20: Maintenance Overview

    Maintenance & Care To assemble the chuck, follow the index finger. There should be only a steps in reverse order. Make sure very thin film of oil left on the chuck. that the chuck is opened during the Please act according to the saying: assembly (the handpiece button Less is more.
  • Page 21: Handling Instructions

    Working with high pressure materials used will not improve the outcome hadewe prevent a formation of your work, since the mo- of rust. In cases in which a tor will even be slowed down. A rust infestation was found within the...
  • Page 22: Troubleshooting

    Use a different bur/tool to test structions in the manual. this. The bur/tool itself can be tested with the hadewe Bur Rubber ring under the knob Shaft Tester (art. 4990) to twists while using it. → Twist see if it is bent.
  • Page 23 Troubleshooting After lengthy use unit does Acceleration spray not spray. stream. Unit is overheated. → Slots of Note: To reduce the starting the unit should not be cov- time of the spray, turn the ered or blocked. Let cool spray regulator on Max. After down the unit for a while.
  • Page 24: Technical Details

    Technical Details 0586 Hyperios Plus Ambient Conditions total weight: 3,5 kg The device is intended for use in dimension: W245xH133xD190 mm dry closed rooms. power supply: 230 V~, 50 Hz rated current: 0,5 A Operation: fuse primary: Temp.: +10°C to +26°C F1, F2: 500 mA delay action fuse Rel.
  • Page 25 For disposal of this unit out tive parts. of Germany, please contact the site Necessary function inspection: where you bought this hadewe speed, display, on/off, spray, clamp- product. ing system. Inspection before use During the production control the following inspections as per DIN EN...
  • Page 26 Technical Details Symbols m1, m2 memory buttons for speed Washer change is necessary. Caution! confirm Important Note! alternating current Follow instructions for use! direct current safety fuse temperature limit (indicates the lower and upper tempera- type B applied part ture limit) socket foot switch air pressure: permissible range date of manufacture,...
  • Page 27 Technical Details Information on electromagnetic compatibility in accordance with DIN EN 60601-1-2 Guidance and Manufacturer's Declaration – Electromagnetic Emissions The device is suitable for use in the specified electromagnetic environment. The customer and/or the user of the pa-on should assure that it is used in an electromagnetic environment as described below: Emissions Test Compliance...
  • Page 28 Technical Details Guidance and Manufacturer's Declaration – Electromagnetic Immunity IEC 60601-1-2 The device is suitable for use in the specified electromagnetic environment. The customer and/or the user of the device should assure that it is used in an electromagnetic environment as described below: Immunity Test IEC 60601 Test...
  • Page 29 Technical Details Guidance and Manufacturer's Declaration – Electromagnetic Immunity IEC 60601-1-2 The device is suitable for use in the specified electromagnetic environment. The customer and/or the user of the device should assure that it is used in an electromagnetic environment as described below: Immunity Test IEC 60601 Test Level...
  • Page 30 Technical Details Recommended Separation Distances between Portable and Mobile RF Communications Equipment and the pa-on Parometer IEC 60601-2 The device is intended for use in an electromagnetic environment in which radiated RF dis- turbances are controlled. The customer or the user of the device can help prevent electromag- netic interference by maintaining a minimum distance between portable and mobile RF com- munications equipment (transmitters) and the pa-on Parometer as recommended below, ac- cording to the maximum output power of the communications equipment.