If the connection inside the fuse is not broken, the issue is with the JACO Power Inverter
assembly and you will need to contact JACO Customer Support to address replacement.
Please see page 3 for locating the correct information that will be needed to start the
replacement process.
If one or both fuses show a break, they will need to be replaced. Contact JACO Customer
Support for details.
Power system is providing power to the LCD and PC, but I do not
have any front status LED indication showing.
1 – Check the Status LED Cable making sure that it is connected properly. You will need to
remove the protective cover inside the Cart top assembly. First remove the Cart work
surface and then, using an 11/32" nut driver, remove the protective cover shown below.
If not seated correctly, position and reconnect. If the front status is still not working, check
the other end of the cable connection. Follow the cable from the front status enclosure and
21