Section 9 Troubleshooting - Hach TitraLab AT1102 User Manual

Hide thumbs Also See for TitraLab AT1102:
Table of Contents

Advertisement

Section 9
Error/Warning
Measurement unstable
Measurement out of range
Temperature out of range
Max. ordinate reached
Below minimum range
Above maximum range
Insufficient curve resolution
Not enough excess titrant
Too much excess titrant
Flat signal
Analysis skipped
Troubleshooting
Refer to the following tables for problem messages and possible causes and solutions.
Table 9 Measurement
Description
The measurement from the sensor is not
stable
The sample concentration is outside the
sensor measurement limits
The sample temperature is outside the
sensor measurement limits
The titration is stopped because the
measured signal is above or below the
ordinate stop point set for this application,
depending on the curve direction
This message is specific to amperometric
applications; the equivalence point has not
been found
This message is specific to amperometric
applications; the equivalence point has not
been found
This message is specific to amperometric
applications; the equivalence point has not
been found
This message is specific to amperometric
applications with back titration; the
equivalence point has not been found
This message is specific to amperometric
applications with back titration; the
equivalence point has not been found
The titration is stopped because the amount
of sample is below the detectable limit of the
application
The titration has been stopped manually by
the user; the equivalence point has not been
found
Possible solution
Make sure that there are no bubbles around the
sensor element or junction
Adjust the sample preparation to be inside the
sensor measurement range
Use a different sensor with a more applicable
measurement range
Adjust the sample temperature to be inside the
sensor measurement range
Use a different sensor with a more applicable
measurement range
Make sure that the sensor is operating correctly
Refer to the application note for best practices
and do the titration again
Adjust the maximum ordinate for the application
(refer to
Change application parameters
on page 63)
Select an application range more adapted to the
sample
Select an application range more adapted to the
sample
Refer to the application note for best practices
and do the titration again
Clean the sensor using the recommended
procedures in the sensor documentation
Refer to the application note for best practices
and do the titration again
Check the excess titrant defined in the
application file and adjust if necessary or select
an application range more adapted to the
sample
Refer to the application note for best practices
and do the titration again
Check the excess titrant defined in the
application file and adjust if necessary or select
an application range more adapted to the
sample
Refer to the application note for best practices
and do the titration again
Select an application range more adapted to the
sample
Preferably use an automatic end titration
detection instead of a manual skip
49