Troubleshooting - Hach TitraLab AT1102 Basic User Manual

Titralab at1000 series
Hide thumbs Also See for TitraLab AT1102:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
Other maintenance options
1. Select an option, then push Select.
Option
Stirring activation
Live measure
Maintenance
summary
Maintenance
schedule
Reagent
replacement

Troubleshooting

Refer to the following table for common problem messages or symptoms and possible causes.
Error/Warning
Stop requested
Measurement unstable
Measurement out of range
Temperature out of range
Slope out of range
Offset out of range
Cell constant out of range
Equiv. point not found
22 English
Description
Do a check of the magnetic stirrer. Push the up and down arrow keys to increase or
decrease the stirring speed.
This option is not available for all sensors. The option shows continuous
measurement data with connected sensors to quickly check measurements. The
installed applications and the automatic additions to the sample are set to off.
Continuous measurements are not temperature compensated, so measurement
differences may occur in the same sample between continuous measurements and
measurements that use installed applications with temperature compensation.
See the number of days remaining for maintenance tasks. After doing a task, push
Reset to set the number of days remaining to the default value.
See the list of maintenance tasks. Push Edit to change the default value.
Use this option to replace the reagents. Obey the instructions on the display.
Description
The operator has pushed the stop button
The measurement is unstable
The measurement is out of range
The temperature measurement is out of range
The calibration slope is outside accepted limits
The calibration offset is outside accepted limits
The calibration value of the cell constant is outside accepted
limits
The titration fails to determine the equivalence point