Change application parameters
Parameter
Type
Part of name
Recommended electrode
Calibration frequency
Stability criterion
Max. stability time
Stirring speed
Stirring duration
Max. temp. variation
Min. slope
Max. slope
Min. offset
Max. offset
Calibration mode
Buffer set
Number fixed buffers
Fixed buffer #n
Calibration mode
Standard
Min. cell constant
Max. cell constant
Temp. compensation
Linear temp. coeff.
Reference temp.
Working mode
Current (sample analysis) Set the value of the imposed current during sample analysis
Current (titrant calibration) Set the value of the imposed current during titrant calibration
Voltage (sample analysis) Set the value of the imposed voltage during sample analysis
Voltage (titrant calibration) Set the value of the imposed voltage during titrant calibration
64
Table 22 Electrode
Description
Type of measurement electrode
Part of the name used to identify matching electrodes if the
preferred electrode is not connected
Recommended electrode to be used for this analysis
Recommended frequency of calibration (0 = no reminder)
Refer to the working procedure for the recommended value
Maximum time limit for the stability check
Speed of the stirrer during electrode calibration
pH electrode parameters
Stirring duration before buffer measurement (0 = buffer
measurement with stirring)
Maximum permitted temperature variation during calibration
Minimum permitted slope limit for the calibration result
Maximum permitted slope limit for the calibration result
Minimum permitted offset limit for the calibration result
Maximum permitted offset limit for the calibration result
Set the electrode calibration mode
List of available buffers for electrode calibration
The number of fixed buffers to use
Buffer to be used during calibration
Conductivity electrode parameters
Set the electrode calibration mode
Standard used for electrode calibration
Minimum permitted cell constant for the calibration result
Maximum permitted cell constant for the calibration result
Follow Natural Water Standards ISO/DIS_7888
Coefficient for linear temperature compensation mode
Reference temperature for linear temperature compensation mode
PtPt electrode parameters
Specify if the probe is used in imposed voltage/current (AC or DC)
Basic mode Expert mode
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x