Maintenance

8.7.5 Pump cassette replacement

44
3. Push the left or right arrow key to select the expert mode, then push Done.
4. From the home screen, select Maintenance, then push Select.
5. From the maintenance screen, select Syringe verification, then push Select.
Note: When the option is selected for the first time, enter the license key.
6. Push the up and down button to select and change the syringe test options.
Option
Description
Syringe
Selects the syringe. Used when there are two syringes.
Shows the syringe information: syringe number and volume.
Operator
Selects the operator ID
Configuration Sets the test settings: test volume and number of replicates. The total
dispensed liquid for the test shows when the two parameters are set.
Note: Make sure to use the applicable beaker and balance for the test based
on the amount of total dispensed liquid.
A field to write the test conditions is available in the test conditions settings.
Start
Starts the syringe verification procedure.
7. Push Start to start the syringe verification procedure. Follow the procedure supplied
with the syringe verification kit.
8. To cancel the syringe verification test and not save data to the data log, push Stop.
9. To cancel the syringe verification test and save the syringe verification results for fully
tested volumes to the data log, push Skip.
10. When the test is complete Pass or Fail shows. Push the left or right arrow to show the
results for the different volumes. Push the up and down arrow to show more results
data for the selected volume.
Electrocution hazard. Remove power from the instrument before this procedure is started.
This option only applies to instruments with pumps installed.
D A N G E R