Differences Between The M9391 And M9393 - Keysight Technologies M9393A Programming Manual

Pxie performance vector signal analyzer
Hide thumbs Also See for M9393A:
Table of Contents

Advertisement

Differences between the M9391 and M9393

The major differences between the M9391 and M9393 are as follows:
M9393 supports noise correction, M9391 does not.
M9393 supports digital image rejection, M9391 does not.
M9391 has a dual-downconversion path across most of the band, M9393
only has this in the narrowband.
M9393 has a Stepped Spectrum IVI interface, M9391 does not.
M9393 has a MultiAcquisition mode in addition to the List mode, M9391
does not.
M9393 has two IF filter bandwidths, where M9391 has only one.
M9393 has three different offset adjustments: IF, digitizer, and mixer,
whereas M9391 only has two.
The interface for alignments is slightly different: there are more options with
M9393. Additionally, the alignments are stored in EEPROM, alignments
therefore don't need to be re-run in between IVI sessions.
M9393 has a Live SFP, M9391 does not.
M9393 requires that the RF input have no incident signal when performing
alignments, M9391 does not.
M9393 has a broadband trigger in addition to the other triggers present for
the M9391.
Keysight M9393A PXIe Performance Vector Signal Analyzer
66

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the M9393A and is the answer not in the manual?

Table of Contents