Summary of Contents for Four E's Scientific CF0201002
Page 1
High Speed Micro Centrifuge Operating Instructions Safety Instructions ................2 Installation ..................2 Structure ................... 3 Operation Panel ................3 Rotor Loading ................... 4 Operation ..................4 Maintenance ..................5 Trouble Shooting ................5 Open the door ................... 6 Accessory ..................6 Specifications ...................
Page 2
Safety Instructions strictly prohibited. • The machine must be placed on a firm and flat table. • For your protection Make sure the centrifuge is horizontal before running. • Make sure the angle between the door and the cover is bigger than 70 degrees when the door is open.
Page 3
• Ideal ambient temperature is 20°C ± 5°C. Do not place the Operation Panel centrifuge where the temperature exceeds 30°C. • To avoid electrical shocks, ensure your hands are dry when Pulse touching the power cord. Door • Make sure the centrifuge to be grounded properly. Structure View Window Prog...
Page 4
Rotor Loading Operation (1) Set the speed RPM/RCF • Load the supplied rotor onto the shaft and make sure the rotor is in position and connected well with the shaft. • Press the speed knob once and the speed value will flash •...
Page 5
• detergent solution. After set the time without pressing the button for confirmation, the • system will stop the value flash and confirm the set value after Sterilize the centrifuge with cloth moistened with 70% ethanol several seconds. solution. • Do not directly pour water, neutral detergent or disinfectant solution (3) Running into the rotor chamber, otherwise the liquid would leak into the drive...
Page 6
1* power adapter Use the screw driver into the door Door failure release hole Specifications Open the door Items CF0201002 (1) If powered on 15000rpm (500-15000rpm), Maximum Speed (rpm) increment:100rpm • The door will open automatically after the running completed.
Page 7
• 50Hz/60Hz 2.5A Failures caused by using non-standard spare parts or accessories ≦56 (5000rpm) and unauthorized modification of the rotor or centrifuge Noise level (dB) • Failures caused by natural disasters including fire, earthquake and 30s↑ 10s↓ Acceleration/Braking time [Sec] so on.