Temperature Calibration; Materials Required; Temperature Calibration Procedure - Bacharach MONOXOR PLUS Configuration And Operation Manual

Carbon monoxide analyzer
Hide thumbs Also See for MONOXOR PLUS:
Table of Contents

Advertisement

Find Quality Products Online at:
Calibration and Maintenance
Step
B-SMART
NOTE: If the correct code was entered, the analyzer accepts it
and returns to the CALIBRATION MENU. If an incorrect code
was entered, the screen will display "Invalid Code." Check to
make sure the correct code has been entered. If problem
persists, contact your nearest Bacharach Service Provider.
NOTE: B-SMART
User Software (FUS).
NOTE: Installing a B-SMART
zero function (either manual or automatic).
NOTE: Bacharach offers a convenient Exchange
Program (where available) that allows the
customer to regularly receive pre-calibrated
replacement sensors that include a code that can
be entered into the analyzer for a quick convenient
setup. Contact Bacharach customer service for
more details about this program.

5.5. Temperature Calibration

This procedure first zeroes and then spans the temperature channel to known
temperature values.
The use of an electronic thermocouple simulator is the preferred method of
producing the desired calibration temperatures. Alternatively, ice and boiling
water baths can be used.

5.5.1. Materials Required

Thermocouple simulator (K-type)
Range:
0 to 600° F (-18 to 316° F)
Accuracy:
(Alternatively) ice water, boiling water, thermometer

5.5.2. Temperature Calibration Procedure

44
GlobalTestSupply
www.
Monoxor
CO Sensor Replacement
codes can be entered through the Fyrite
sensor forces the instrument to perform a
± 0.5° F (± 0.3° C)
Plus Manual
0019-9370 Rev 1
.com