5.2 Ambient Temperature Calibration
Material Required:
Procedure:
1. Enter the Calibration Mode as described in Section 5.1. Then re-
peatedly press the ENTER button until "AMBIENT" appears at the
top of the display.
2. Use the L and M buttons to set the displayed value to match the
reading of a calibrated thermometer at room temperature.
3. Press ENTER to store the displayed value and move to the next
calibration screen, or hold down ENTER for 2 seconds to store the
displayed value and enter the Run Mode, or press the I/O button to
exit the Calibration Mode and turn OFF the analyzer without sav-
ing the changes.
5.3 TA (Stack Temperature) Thermocouple
Channel Calibration
Material Required:
Procedure:
1. Attach the thermocouple simulator to the Stack Ther mo cou ple con-
nector on the analyzer (T-STACK, see Figure 3-2).
2. Enter the Calibration Mode as described in Section 5.1. Then re-
peatedly press the ENTER button until "TA" appears in the top left
side of the display.
22
Fyrite Tech
Calibrated Thermometer
Thermocouple Simulator
Instruction 24-9435