Section 7: Scanning resistors using 4W measurement
DAQ6510 Data Acquisition / Multimeter System User's Manual
To highlight the impact of test lead resistance on low-resistance measurements, and how the
correction methods improve readings, the connections to each DUT consist of one meter of 22 AWG
wire.
To prevent electric shock, test connections must be configured such that the user cannot
come in contact with test leads or any device under test (DUT) that is in contact with the
conductors. It is good practice to disconnect DUTs from the instrument before powering the
instrument. Safe installation requires proper shields, barriers, and grounding to prevent
contact with test leads.
There is no internal connection between protective earth (safety ground) and the LO
terminals of the DAQ6510. Therefore, hazardous voltages (more than 30 V
) can appear on
rms
LO terminals. This can occur when the instrument is operating in any mode. To prevent
hazardous voltage from appearing on the LO terminals, connect the LO terminal to protective
earth (safety ground) if your application allows it. You can connect the LO terminal to the
chassis ground terminal on the front panel or the chassis ground screw terminal on the rear
panel. Note that the front-panel terminals are isolated from the rear-panel terminals.
Therefore, if you are using the front-panel terminals, ground to the front-panel LO terminal. If
using the rear-panel terminals, ground to the rear panel LO terminal. Failure to follow these
guidelines can result in injury, death, or instrument damage.
7-4
DAQ6510-900-01Rev. A / April 2018