Precision depending on position
(Note Fig. 3)
Measuring position
8. Operating Instructions
General information
Always switch OFF the device when it is not in use.
Please refer to the sketches on the first pages of this manual.
To switch the unit on or off, press the main switch (5) until a short beep is
heard.
Attention!
Avoid voltage measuring in electrical circuits while motors are switched on or
off. The stress-spikes can damage the instrument.
Hazardous voltage! The probes may not be long enough to touch the hot parts
in some 230V wall sockets as they are deep inside. As a result, the reading can
show 0 volts. Make sure that the probes touch the metallic contacts in the
socket before assuming that voltage has not been applied.
Devices like welding transformer, car ignition system, etc. could produce stray
electromagnetic fields which could adulterate the result of a measurement.
A
35 mm
B
50 mm
C
60 mm
D
>60 mm
EN 8
Deviation
±0.5%
±1.5%
±2.0%
±5.0%