Test Measurement; Adjustment - Test Equipment Depot SBS 2500 Manual

Table of Contents

Advertisement

4.2

Test measurement

- Push drain button (3) downward completely.
- Ensure that the sample tube is located in the accompanying density standard
or in distilled water.
- Press the fill button (12) slowly to fill the measuring cell.
- Ensure that no air bubbles are contained within the measuring cell.
The instrument automatically executes a measurement and the result appears in
the display.
The result for water must be 1.000 in the selected measurement unit SG (t/t).
If the deviation is ≤ 0.0005
Adjustment is ok, the instrument is ready for measurement.
If the deviation is > 0.0005
Adjust instrument, see Chapter 4.4.
4.3

Adjustment

Warning: you must be in the "expert" mode to perform a water calibration
Initializing adjustment
- Press and hold the Í key until
The instrument adjusts automatically (duration: approx. 1 minute). After adjustment
is completed, the measured deviation from the theoretical value and Execute?
(
) appears.
No
- Press the È or Í key. Execute? (
- Press the ok/meas. key to confirm.
The adjustment is confirmed.
6
3
12
CALIB (Water)
appears on the display.
) appears.
Yes