Replacing The Sample Pump; Standard And Optional Equipment; Standard Equipment; Optional Equipment - Test Equipment Depot SBS 2500 Manual

Table of Contents

Advertisement

Cleaning / Standard and optional equipment
8.4

Replacing the sample pump

Removing cover of sample pump
- Slightly press together cover of sample pump on grip recesses and detach it.
- Pull sample pump out of guide system.
Inserting sample pump
- Insert new sample pump into guide system.
Ensure that the pump plunger (2) and pump body (3) are situated in the cor-
responding guides when inserting the sample pump.
- Check for proper functioning of sample pump.
- Put the cover of the sample pump back in its place.
9

Standard and optional equipment

Each part identified by an order number can be ordered from SBS.
9.1

Standard equipment

The instrument is delivered pre-assembled.
1
SBS 2500 Specific Gravity Meter
with carrying case, includes:
2
AAA-type batteries (LR03, 1.5 V)
1
Operating instructions
9.2

Optional equipment

Sample tube (with washer)
Sample pump, compl.
O-ring for sample pump
SBS 2500
Order No.
SBS 2500
51710073
Order No.
51324641
51324405
51524506
Operating Instructions
Bedienungsanleitung
Mode d'emploi
Manual de Instrucciones
Istruzioni d'uso
Densito 30P
15