Measurement (meas)
5.4

Measuring

The
symbol appears in the display: Measuring with automatic stability
control
- Press the ok/meas. key to start measuring.
The
symbol blinks during measurement. The result appears in reverse display.
The
symbol does not appear on the display: Measuring with manual stability
control
The instrument measures continuously.
- Wait until the displayed value stabilizes.
- If the result is to be saved: Press the ok/meas. key.
The result appears in reverse display.
Emptying the measuring cell
- Hold the sample tube over a waste container.
- Press drain button (3) downward slowly and move it up again with the fill button
(12).
5.5

Saving the results

The instrument can save up to 1100 results internally. Each time a result is saved,
the number of internally-stored values increases by one.
The procedure for saving the results is dependent upon the settings in the menu
(see Chapter 5.4).
The
symbol appears in the display: Saving all results automatically
The instrument saves all results automatically.
The
symbol does not appear in the display: Saving selected results
manually
Save the result:
- Press the ok/meas. key.
Do not save the result:
- Press the esc key.
5.6

Displaying and marking saved results

Displaying saved results
- Press the Á key.
The sample number flashes and the symbol
- Scroll through the saved results using the È and Í keys.
Marking results
For the identification of invalid or incorrect results or for marking a sample
change.
- Select the desired sample number using the È and Í keys.
- Press the ok/meas. key.
The selected sample number is marked with an asterisk.
Note
The marking is removed if the sample has already been marked.