Trouble Shooting - BRAND Transferpette Series Operating Manual

Table of Contents

Advertisement

Trouble Shooting

Problem
Pipette tip leaks or
volume too low
Poor sample aspiration
Volume too large
Pipetting key moving
stiffly
Possible Cause
- Tip does not sit properly
- Damaged V-ring (O-ring) on single
nose cone
- Piston damaged
- Seal dirty, damaged
- Instrument is dirty
- Non-aqueous solutions pipetted
- Single nose cone clogged
- Pipetting key pressed too far down
into the blow-out stroke before
aspirating liquid.
- Key movement reduced
- 32 -
Corrective Action
- Press tip on firmly. Use only suitable
high-quality tips, page 8.
- Replace V-ring (O-ring) on nose cones.
"Repairs", pages 26-28.
- Replace piston.
"Repairs", pages 24, 25.
- "Repairs", pages 22, 23.
- Clean instrument.
"Cleaning", page 19.
- "Adjustment", page 14.
- Clean nose cone. Remove cones only;
do not disassemble instrument!
"Cleaning" page 19.
- Operate pipette properly.
"Operating the pipette", page 10.
- Lubricate piston and centrale guide rod
with the appropriate grease,
pages 13, 34.