8.9

Messages about EAM exhaust-gas analyzer module

No.
Message
4000 Error opening interface
4001 Error writing to interface
4002 Error reading interface
4003 Error closing interface
4004 Exhaust-gas interface not open
4005 Command not allowed at present time
4006 Unexpected / unknown reply
4007 Memory full
4008 Exhaust-gas analyzer module reports
parameter error
4027 Calibration of HC channel outside tolerance
4028 Calibration of CO channel outside tolerance Replacing measuring bank
4029 Calibration of CO
4030 O
sensor voltage below minimum
2
allowable level
4031 Supply voltage outside tolerance
4032 Analyzer part temperature measurement
malfunction
4033 Faulty air-pressure measurement
4034 Insufficient flow
4035 No temperature compensation
4036 Recalibrate with calibrating gas
4040 Calibration for HC signal outside tolerance
4041 Calibration for CO signal outside tolerance
4042 Calibration for CO
4043 Analyzer part does not respond
4044 Incorrect channel assignment
4050 HC channel not calibrated
4051 CO channel not calibrated
4052 CO
channel not calibrated
2
4053 Carbon canister contaminated with
HC residues
4055 Leak test failed
4057 HC residues in gas sampling system
channel outside tolerance
2
signal outside tolerance
2
Remedy
Software error: Reboot system, if error still present:
Re-copy system software on VSM, replace control
module PCB
Replace O
measuring sensor
2
Check continuity of connection cable, replace if necessary
Check mains voltage, if mains voltage is out of tolerance,
changeover transformer connections accordingly.
Check interconnecting cable to DMM, replace if necessary
Check voltage supply
Replace control module PCB
Replace measuring bank
Replace measuring bank
Replace measuring bank
Check coarse filters GF1, GF2, GF3 and GF4,
replace if necessary
Check suction power of pump, replace if necessary
Check flow sensor, replace if necessary
Check hosing as per gas circuit diagram
Clean condensate separator
Replace measuring bank
Perform recalibration with calibrating gas
Replace measuring bank
Replace carbon canister
Check cross sensitivity to water vapour
Replace hosing as per gas circuit diagram
Clean analyzer chambers, not in warranty cases!
Replace measuring bank
Check exhaust-sample probe and hose for cracks
Check hosing as per gas circuit diagram
Perform leak test of measuring bank, replace
measuring bank if necessary
Purge exhaust-gas analyzer module with fresh air,
if error still present: replace internal hosing (see gas
circuit diagram)
Sectio
5.23
10.27
10.10
10.11
12.1 and 12.4
10.1
10.2
10.3
10.27
10.10
10.10
10.10
10.13 u. 10.14
10.15
12.5
10.18
10.10
4.7.5
10.10
10.12
12.5
10.9
10.10
12.5
10.7
10.10
4.7.1
12.5