Retrofit Procedures with fluid in the tank ......................6 Retrieve the TD80 strap table ....................... 6 Exchange the Transmitter ........................6 Transferring the TD80 strap table to the TD100 ................... 6 Performing Offset Calibration on the TD100 ..................7 Confirming Offset Calibration on the TD100 ..................7 Retrofit Procedures with empty tank ........................
When speaking with your Titan Logix representative, please indicate whether the item is still under warranty. Shipping cost to and from a Titan Logix service facility is the responsibility of the customer unless the product is under warranty. If the product is under warranty, the freight cost to Titan Logix is incurred by the customer. From Titan Logix, the cost is incurred by Titan Logix.
Application Note 3. Introduction This document describes the steps that should be followed to ensure a successful replacement of the TD80 transmitter with a TD100 transmitter where there may or may not be fluid in the tank. This includes copying the strapping table and calibration data from the TD80 to the TD100.
4.2 Initial Steps Prior to removing the TD80, determine if fluid exists in the tank, and if so, to what level. If the level of fluid in the tank is between 25cm, approximately 10 inches, above the top of the shorting block and 75cm approximately 30 inches, below the fiducial proceed to section “Retrofit Procedures with fluid in the...
5. Save the TD80 strap table as a file for future reference. 5.2 Exchange the Transmitter 1. Power down the TD80 and Finch II and proceed to replace the TD80 with TD100. 2. Ensure proper electrical connections to the TD100 have been made.
5.4. 5.4 Performing Offset Calibration on the TD100 1. At this point it is assumed the TD80 transmitter has been removed and replaced with a TD100 transmitter. 2. Apply power to the Finch II / TD100 system. DO NOT INITIATE OFFSET CALIBRATION YET.
2. Take note if there is a sump and/or riser on the tank. Record their corresponding values. 3. Read the strap table off the TD80 using SensorLink utility program. a. Information relating to the use of the SensorLink utility program can be found in TPM 058 SensorLink User Manual.
2. Remove the TD80 transmitter from existing installation and connect it to the external Dual-Rod probe with all electrical cabling in place. 3. Power up the TD80 transmitter and record the volume as shown on the Finch II. Note that any objects close to the probe may affect the reading.
TD100. d. Under Step 3: i. Adjust the “HH level” threshold accordingly bearing in mind that the TD80 and TD100 have different minimum distance requirements between HH and Spill threshold. ii. Adjust the “Display Precision” as necessary.
Application Note 10. The TD100 should now contain the same strapping table as stored in the TD80 and an Offset Calibration has been applied. 6.5 Installing the TD100 1. Detach the TD100 from the external Dual-Rod probe and install the TD100 onto the truck.