Overwrite Stored Calibration Data; Retrieve The Factory Calibration - Hach POCKET COLORIMETER II LCK 360 Zinc Application Instructions

Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
System start up
3.
"RES" is displayed. Press BLUE key to scroll through the entries. Compare the
entries with the figures in the
In case a value was not stored correctly, retrieve the factory calibration first. See
Retrieve the factory calibration on page
start the programming procedure again.

3.5 Overwrite stored calibration data

With the POCKET COLORIMETER II it is possible to overwrite stored calibration
data. Before overwriting calibration data it is recommended to retrieve the factory
calibration first. After retrieving the factory calibration, you can program new
calibration data.
In case of any mistake retrieve the factory calibration first.
3.5.1

Retrieve the factory calibration

1.
Press MENU. "SEL" is displayed. A flashing arrow indicates the current
measuring range (1 or 2). Press GREEN key to toggle between measuring
ranges. Select measuring range retrieving the factory calibration.
2.
Press MENU for about five seconds until "USER" followed by "CAL" is
displayed.
3.
Press BLUE key to scroll to "dFL". Press GREEN key.
3.2.2.1 Evaluation as Zinc on page
25. After retrieving the factory calibration,
22.
3.5.1
25

Advertisement

Table of Contents
loading

Table of Contents