"BATT". If this happens, either recharge the battery or select a lower rate to complete the
feed cycle.
• The lifetime of the battery is optimally maintained by using the pump on a regular basis on
battery until it is fully discharged followed by a full recharge via the adapter on mains power.
• In case of battery failure, caused by, for example, an excessive temperature condition (which
may resolve itself with time) or by a complete battery failure, the E, F and battery symbol will
flash. Turn the pump "off" and contact your healthcare provider or refer the pump to service.
Connection to an external alarm system
The power connector at the side of the pump offers the option of connecting the pump via
the Flocare
®
Infinity
Nurse Call (corporate code 35752) to an external alarm system.
This external alarm system may be with open or closed contact.
• Follow the instructions for use of the Flocare

ALARM FUNCTIONS AND SAFETY FEATURES

In case any of the problems listed on the problem solver chart occur, the pump delivers an audible
and visual alarm and stops working. The back light of the LCD automatically switches on.
Exception to this is the low battery alarm "BATT", in this situation the pump continues working.
Action in the event of an alarm:
Check the type of alarm displayed by the large display characters.
• Press the "START/STOP" key to stop the audible and visual alarm.
• Correct the cause of the alarm as described in the table.
• Start the pump again by pressing "START/STOP".
The problem solver chart on the next page gives a clear explanation of the alarms.

MAINTENANCE

Cleaning:
• Always unplug the pump prior to cleaning to avoid electric shock hazard.
• On a regular basis thoroughly clean all surfaces of the pump (including the sensors and rotor)
with warm soapy water, a 5% bleach solution in water, or a multipurpose disinfectant cleaner.
• The Flocare
®
Infinity
pump may be rinsed by holding under a stream of warm, clean water.
Do not submerge the pump!
• Always maintain the three rollers on the rotor in a clean state to ensure they spin freely.
• The adaptor normally does not require cleaning. When desired, a dry or slightly damp cloth
may be used to clean the outside surface of the adaptor. Make sure the adaptor is
disconnected from the wall outlet.
Safety notes:
• The manufacturer recommends an inspection of the pump at an authorised service centre
every 2 years.
• If any fault occurs during use, or if the pump is dropped, it should be checked by authorised
technical personnel prior to use.
UK
• Replace the Flocare
®
Infinity
prevent the growth of harmful bacteria. Dispose of Flocare
properly, as required by local law.
-60-
®
Infinity
Nurse Call to connect the pump.
pump set every 24 hours to maintain delivery accuracy and
®
Infinity
disposable sets

PROBLEM SOLVER CHART

Condition
Cause
NO SET
• The set is not fitted or wrongly
• Stop the alarm by pressing "START/STOP"
fitted in the pump.
• Insert the Flocare
• Restart the pump.
• The pressure sensor area is
• Clean the sensors, reinsert the cassette in the pump
dirty
PUSH STRT
• The pump has been untouched
• Stop the alarm and prolong the hold mode with
in hold mode for 3 minutes or
more.
• Program the pump and start it by pressing the
END OF DOSE
• The pump administered the
• Turn the pump off by pressing the "ON/OFF" key
installed dose = volume.
• Clear the memory of the total volume administered
AIR
• The air sensor detected an
• Replace the empty feeding reservoir and continue
empty pump set.
• The air sensor area is dirty.
• Make sure the air sensor is clean.
• The feeding set is not inserted
• Make sure the feeding set is properly inserted
correctly.
• The feeding set is used
• Replace the feeding set by a new one.
for more than 24 hours.
BATT
• The battery capacity is too low.
• Connect the adaptor to the pump and mains and
the pump
The pump is not able to deliver
remains working
highly viscous fluids at high flow
rates with the present charge
level of the battery.
Battery, "E" and
• Battery failure.
• Turn the pump "off", contact your healthcare provider
"F" flashing
FILL SET
• The pump is priming the set.
• Press the "FILL SET" key another time to stop the
OCC IN
• The pump detected an upstream
• Stop the alarm by pressing the "START/STOP" key.
occlusion between the pump
• Remove the feeding set out of the pump and check
and the feeding bag.
• Re-insert the feeding set in the pump and restart.
• The pressure sensor area is
• Clean the sensors, reinsert the cassette in the pump
dirty.
OCC OUT
• The pump detected a
• Stop the alarm by pressing the "START/STOP" key.
downstream occlusion between
• Check the permeability of the feeding tube by
the pump and the patient.
• The pressure sensor area is
• Clean the sensors, reinsert the cassette in the pump
dirty.
OCC OUT
• Calibration not yet completed
• Stop the alarm by pressing the "START/STOP" key.
(repeated)
on the current feeding set
• Start the pump by pressing the "START/STOP" key
• Stop the pump by pressing the "START/STOP" key
• Remove the feeding set from the pump and reinsert
• Restart the pump by pressing the "START/STOP" key.
LOCK
• Only the current feeding
• The programming feature is blocked in the set up
schedule is allowed for this
patient. Another feeding pro-
gram is not allowed by your
healthcare professional.
Correction
®
Infinity
cassette into the pump as
indicated on the blister packaging and close the door.
and restart the pump.
another 3 minutes by pressing "START/STOP" key.
"START/STOP".
and hold it during 2 seconds or
(see paragraph: clearing the memory), reprogram a
new feeding schedule and start the pump by pressing
"START/STOP".
feeding. If necessary prime the set.
in the pump.
charge the pump for approx. 6 hours. During
charging the pump can be used.
or refer the pump to service.
pump and bring it back in the hold mode.
the permeability by flushing the line.
and restart the pump.
aspirating liquid via the medication port.
and restart the pump.
and let it run only briefly.
insuring that there has been no occlusion out alarm.
the feeding set in the pump.
mode of the pump. Ask your healthcare professional
to modify this setting.
UK
-61-