TPC L700 Lucent Dental Light Operation And Service Manual

Advance operatory dental light

Advertisement

Quick Links

Advance Operatory
Dental Light
Operation and service manual
TPC
851 S. Lawson St
City Of Industry, CA 91748
P626-810-4337 Fax 626-810-4245
www.tpcdental.com
1
REV. 1-1-13

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the L700 Lucent Dental Light and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for TPC L700 Lucent Dental Light

  • Page 1 Advance Operatory Dental Light Operation and service manual 851 S. Lawson St City Of Industry, CA 91748 P626-810-4337 Fax 626-810-4245 www.tpcdental.com REV. 1-1-13...
  • Page 2: Table Of Contents

      Table on Contents:                Page  GENERAL INFORMATION              3        Installation                4    Overview of components          5         Operation instructions     ...
  • Page 3: General Information

      Gene ral Info ormatio n:   Defin itions o of symb ols    Risk of ele ectrical shock  present. Mak ke  Type B equi pment   sure powe er is disconne ected before  (Protected a against electr rical shock)   attemptin g this proced ure.            Caution: F Failure to care efully follow  Dangerous V Voltage  the descri bed procedu re may result t in  damage to o the equipm ment or ...
  • Page 4: Installation

      Operatory Light Installation  • The Advance Operatory Light is designed to fit into the top end of a chair or floor  mounted post. The size of the included bushing is 1 11/16”. For all other light  post sizes you will need to obtain an additional light post adaptor bushing.     • Insert the end of wire through the silver light bushing.       • Place the end of the wire through the light post and insert the light bushing into  the light post and press down to insure proper seating of the bushing.     • Attach light head wires to vertical arm this is a “white Molex connection”. Insert  set screw into light head to attach light to vertical arm.     • Connect the wires to the corresponding terminals on the plug. Or hard wire the  connections using the included wire nuts.                4   ...
  • Page 5: Overview Of Components

      Overview of components:     Vertical Arm   Transformer lid  Horizontal Arm   Light  Knuckle   Transformer    Housing  Glass  Light Handle  Reflector  Light bushing    Deflector   Power Switch  Light Post            5   ...
  • Page 6: Operation Instructions

      Operating Instructions  • Move power switch to the left to set the light to high intensity. Move the power  switch back to the center position to turn it to the off position. Move the power  switch to the right to right to turn it to low intensity.   High Low  Off    • To adjust the position of the light, place your hand on the light handle and move  it to the desired location. You can move the light head up and down and left and  right. When turning the light head left and right it will stop, don’t force any  further movement of the light head as damage may occur.     • To bring the light closer to the patient or desired illuminated location pull down  on the light handle to move the vertical arm. Once the light is in the desired  location I will stay in that positions until you move it back.         6   ...
  • Page 7: Light Head Adjustment

      Light head adjustment  • If you experience a downward drift with your operatory light head please follow the  procedure below.   • Tighten both set screws on the back side of the operatory light. There is one located on  each side of the dental light head. Reference image below. To tighten the set screw turn  it clockwise. Do not over tighten the set screw as damage may occur.                   Vertical Arm adjustment  • If you experience a downward drift with your operatory lights vertical arm please follow  the producer below.   • With one hand lower the vertical arm till you see the adjustment. See image below.  Adjust the inner nut till the desired tension is achieved.       Less tension More tension Turn tension nut counter  Turn tension nut  clockwise if vertical light  clockwise if vertical light  arm drifts up.   arm drifts down.   7   ...
  • Page 8: Replacement Of Light Bulb

      Light Bulb replacement • If you need to replace your operatory light bulb please follow the steps listed below. • Remove the rear panel cover from the back of the light. Turn the thumb nut counter clockwise to remove it. Thumb Screw  ...
  • Page 9: Replacement Parts List

      Replacement Parts List Index No. Part Number Description L700-0001 Housing cover for transformer P56004-700 110V Power Transformer L700-0002 Transformer Housing L700-0003 Horizontal Arm L700-0004 Counter Sunk Pin L700-0005 Swivel Arm Knuckle L700-0006 Bridge ARM Connector L700-0007 Set Screw (4 X 5 mm) L700-0008 Yoke Joint Cover L700-0009...
  • Page 10   10   ...
  • Page 11   Index No. Part Number Description L700-703 Light Shield L700-702 Light Handle L700-701 Light Head Yoke L700-704 Frame L700-705 Yoke Cover L700-706 Frame Back Cover L700-707 Yoke Wire Harness L700-708 Power Switch L700-709 Joint Nut L700-710 Flat Washer L700-711 Wave Washer L700-712 Nut For Joint Bolt L700-713...
  • Page 12: Cleaning & Disinfecting Guide

      Cleaning and disinfecting guide   Table 1  *The Manufacturer makes no representation as to the disinfectant efficacy of these products.  We make no warranty expressed or implied that these disinfectants will not damage the  surface finishes. Damage and discoloration of the surface finishes are not covered under the  warranty. Iodophor‐based disinfectants will cause yellow staining on many surfaces. Regular  washing with soap and water will minimize this staining. Iodophor neutralizers such as  Promedyne are also available.      Unacceptable Disinfectants  Conditionally Acceptable Disinfectants    These disinfectants will harm the surface  These disinfectants have been found to be the    finishes of dental equipment and are not  least  recommended.   harmful to the equipment surfaces by our test  Use of these products will void your  methods.    warranty.  Chemical Composition Trade Names  Chemical Composition / Trade Names  Iodophors  Biocide, Aseptic‐IDC,  Strong Phenols / Lysol, Lysol 2, Lysol  Wescodyne, SD5, Promedyne,    Phenol‐Alcohol Professional, Coe Foam, Coe  Iodo‐Five  combinations Spray Pump, Vitaphene, Omni II  Mild Phenols Procide ES, Asepti‐Steryl    Sodium Hypochlorite / Clorox, Ajax, Purex  Aerosol  Household Bleach  Glutaraldehyde / Sterall Spray, Coldspor  Alcohol ...
  • Page 13: Warranty

    TPC will repair or replace any defective part at no charge. TPC will not be responsible for labor charges or shipping charges to / from the TPC facility. This guarantee does not cover normal wear or stains on surface finish. The guarantee does not cover damage resulting from improper installation, misuse or accidents incurred in shipping and handling.
  • Page 14: Troubleshooting Guide

      Troubleshooting Guide.   Problem   Cause  Solution   Light does not illuminate   No power to light   Check power     Light bulb has expired   Replace light bulb     Transformer is expired   Replace transformer   Only high or low function  Secondary windings on  Replace transformer   works   transformers are expired     Main power switch is bad   Replace main power switch  Light turns on and off then  Short or open in power line  Check each power line for  the light head is moved   form transformer.   short or open. Then replace.   Light head drifts down   Light head tension springs are  Tighten light head spring nuts.  loose  Vertical arm drifts up or down  Vertical arm tension spring  Adjust vertical arm tension.   needs to be adjusted   Light drifts from left to right   Light post is not level   Level light post.      ...