Verifying the Wideband AC Option

Wideband verification verifies that the 5790B WIDEBAND input (requires Option
5790B/3, 5790B/5, or 5790B/AF) is within tolerance. There are two worksheets
and one test record to facilitate this procedure. You will need 1 copy of Table 3-
15, 8 copies of Table 3-16 (one for each voltage range) and 1 copy of the overall
test record, Table 3-17.
If verifying Option 5790B/3, frequencies above 30 MHz are not
applicable.
Range
7 V
2.2 V
2.2 V
2.2 V
700 mV
220 mV
70 mV
22 mV
7 mV
2.2 mV
7 mV
3-52
Table 3-15. Wideband 1 kHz Gain Verification Worksheet
Nominal @
Measured
1 kHz
by 5790B
3.2 V
2.0 V
1.0 V
0.64 V
320 mV
100 mV
32 mV
10 mV
3.2 mV
1 mV
3.2 mV
Note
5730A Value
(Recorded
5790B Error
in Table 3-8)
Lower Limit
Upper Limit