Secondary Electrical Specifications; Ac Secondary Performance - Fluke 5790B Service Manual

Ac measurement standard
Hide thumbs Also See for 5790B:
Table of Contents

Advertisement

Secondary Electrical Specifications

Secondary performance specifications and operating characteristics are included in uncertainty specifications.
They are provided for special calibration requirements such as stability or linearity tests.

AC Secondary Performance

Voltage
Frequency Range
Range
10 Hz - 20 Hz
20 Hz - 40 Hz
40 Hz - 20 kHz
20 kHz - 50 kHz
2.2 mV
50 kHz - 100 kHz
100 kHz - 300 kHz
300 kHz - 500 kHz
500 kHz - 1 MHz
10 Hz - 20 Hz
20 Hz - 40 Hz
40 Hz - 20 kHz
20 kHz - 50 kHz
7 mV
50 kHz - 100 kHz
100 kHz - 300 kHz
300 kHz - 500 kHz
500 kHz - 1 MHz
10 Hz - 20 Hz
20 Hz - 40 Hz
40 Hz - 20 kHz
20 kHz - 50 kHz
22 mV
50 kHz - 100 kHz
100 kHz - 300 kHz
300 kHz - 500 kHz
500 kHz - 1 MHz
10 Hz - 20 Hz
20 Hz - 40 Hz
40 Hz - 20 kHz
20 kHz - 50 kHz
70 mV
50 kHz - 100 kHz
100 kHz - 300 kHz
300 kHz - 500 kHz
500 kHz - 1 MHz
10 Hz - 20 Hz
20 Hz - 40 Hz
40 Hz - 20 kHz
20 kHz - 50 kHz
220 mV
50 kHz - 100 kHz
100 kHz - 300 kHz
300 kHz - 500 kHz
500 kHz - 1 MHz
Temperature Coefficient
24 Hour AC
Stability ± 1 °C
10 °C to 40 °C
Slow Filter Peak-
Peak ± μV
0.4
50
0.4
50
0.4
50
0.4
50
0.8
75
1.5
100
3.0
150
4.5
200
0.4
15
0.4
15
0.4
15
0.4
15
0.8
25
1.5
60
3.0
80
4.5
125
0.4
0.4
0.4
0.4
0.8
1.5
10
3.0
40
4.5
100
± (ppm of Reading)
18
18
18
18
24
24
10
48
30
150
75
12
1.5
8
1.5
8
1.5
8
2.0
18
5.0
24
10.0
36
20.0
120
50.0
Introduction and Specifications
Secondary Electrical Specifications
[1]
0 °C to 10 °C
40 °C to 50 °C
ppm / °C
50
50
50
50
75
100
150
200
15
15
15
15
25
60
80
125
5
5
5
5
5
5
5
5
8
8
10
40
100
5
5
5
5
5
5
5
5
8
8
10
30
75
3.0
3.0
3.0
3.0
8.0
10.0
20.0
50.0
1
Input
[2]
Resistance
10 MΩ
10 MΩ
10 MΩ
10 MΩ
10 MΩ
1-13