6.9 SWEEP Mode
For 4-pole earth measurements (without clamp, selective with clamp, or with the
GroundFlex
), it is possible to select the frequency sweep mode (SWEEP).
®
Press the Hz/OPTIONS button, then the ► button. The device then displays the
first free memory location (OBJ:TEST) in which to store a series of measurements
with a series of preset frequencies.
The device makes a measurement at each of the following frequencies:
41
64
92
128
256
513 1025 1465 2051 2539 3125 3516 4102 4687 5078
These SWEEP frequencies are factory defaults, preset in the memory. The user
can change them by configuring the tester using DataView
software.
®
6.10 Smoothing
In the manual mode, you can activate or deactivate the smoothing of the measure-
ment results by pressing the 2nd + DISPLAY (SMOOTH) buttons. This smoothing
consists in displaying an exponential mean value, a significant help with highly
fluctuating values.
52