5 Emissivity - Typical Values

5.1 Setting the Emissivity value

By default, the emissivity value is set remotely via the Landscan Configuration Software
or by Landscan WCA software, if this option is used.
For further information, see the relevant User Guide.
Alternatively, the emissivity setting can be set to a fixed value using the procedure
below.
1)
Unscrew the four bolts holding the scanner side plates to the body using a 5mm
allen/hex key.
2)
Notice that the two sides of the scanner are dissimilar and identify the 'open' side
where the external parts of the optical system can be clearly seen.
3)
On the 'open' side of the scanner, find the jumper marked JM3 towards the right
edge of the PCB (shown inside green circle below). Remove the jumper from the
PCB.
LSP
Linescanner Head
HD