Ametek LAND LSP HD Series User Manual page 12

Linescanner head
Table of Contents

Advertisement

LSP
Linescanner Head
HD
1.2.5 LSP
6 series specifications
HD
Operating
Wavelength
Temperature range
System Accuracy
(1)
(5 to 95% of range)
Focus distance
Target width
(static to 95% radiance)
Target distance <1200 mm / 47 in
Target distance 1200 to 3000 mm /
47 to 118 in
Repeatability
Temperature
Resolution
(2)
Below 15% of range
(noise) Above 15% of range
Scan Angle
(3)
Scan Rate
Factory set
User adjustable
Speed of response
(Tau - linear output on rising edge)
Ambient Stability
deltaT[indicated] / deltaT[ambient]
Settling Time
(4)
*LSP
61, LSP
62, and LSP
63 as per LSP
HD
HD
HD
(1) Includes error correction
(2) No line averaging. 4LA reduces noise by factor 2
(3) Scan angle may be set in the range 40 to 80° by the PC software
(4) For optimum performance, the thermoelectric cooler takes time to stabilise the detector thermally. This process takes longer at higher ambient
temperatures.
Page 6
LSP
60
LSP
HD
3 to 5 µm
3 to 5 µm
20 to 250 °C
50 to 400 °C
68 to 482 °F
122 to 752 °F
± 2 °C / 3.6 °F
Fixed at 1200 mm / 47 in
12 mm / 0.5 in
Field of view of 100:1 (0.01 radians)
< 0.5 °C / 0.9°F
≤ 1.5°C / 2.7°F ≤ 1.5°C / 2.7°F ≤ 2.5°C / 4.5°F ≤ 2°C / 3.6°F
≤ 1°C / 1.8°F
≤ 1°C / 1.8°F
80 °
10 to 150Hz in 1 Hz steps
20 Hz
50 Hz
≤ 5µs
< 1° / 10°
< 1° / 10°
2 Hours
60 unless otherwise stated
HD
61*
LSP
62*
HD
HD
3 to 4.2 µm
100 to 600 °C
212 to 1112 °F
≤ 2.5°C / 4.5°F
50 Hz
< 2° / 10°
User Guide
LSP
63*
HD
50 Hz
≤ 10µs
< 1° / 10°

Hide quick links: