Specifications; General Information - Hach 2100AN IS Basic User Manual

Hide thumbs Also See for 2100AN IS:
Table of Contents

Advertisement

Available languages

Available languages

Specification
Details
Stabilization time
Immediately
Reading modes
Manual or auto range, signal averaging on and
adjustable or off, Ratio on or off
Power requirement
115–230 VAC, 50/60 Hz (automatic power selection)
28 W maximum
Pollution
2; II
degree/installation
category
Protection Class
1
Operating conditions
Temperature: 0 to 40 °C (32 to 104 °F)
Relative humidity: 0–90% at 25 °C, 0–75% at 40 °C,
noncondensing
Altitude: 2000 m (6560 ft) maximum
Indoor use only
Storage conditions
–40 to 60 °C (–40 to 140 °F), instrument only
Printer
Built-in (thermal, 58-mm, up to 28 column)
Interface
RS232C serial interface by way of DB9 subminiature
D-shell connector for data output to computer or
printer, and data input (command). No handshaking.
Air purge
Dry nitrogen or instrument grade air (ANSI MC 11.1,
1975)
0.1 scfm at 69 kPa (10 psig); 138 kPa (20 psig)
maximum
Hose barb connection for
Sample cells
Round cells 95 x 25 mm (3.74 x 1 in.) borosilicate glass
with rubber-lined screw caps
Note: Smaller sample cells (less than 25 mm) can be used
when a cell adapter is used.
Sample requirements
25 mm sample cell: 20 mL minimum
0 to 95 °C (32 to 203 °F)
Enclosure
High-impact polycarbonate plastic
4 English
1
/
-inch tubing
8
Specification
Details
Dimensions
30.5 x 40 x 15.6 cm (12.0 x 15.7 x 6.1 in.)
Weight
3.8 kg (8.5 lb)
Certification
CE, cETLus
1
Turbidity specifications identified using recently prepared formazin standard
and matched 25-mm sample cells.
2
Reference conditions: 23 ± 2 °C, 50% ± 10% RH noncondensing,
115/230 VAC, 50/60 Hz
3
Intermittent electromagnetic radiation of 3 volts/meter or greater may cause
slight accuracy shifts.
4
FNU is equivalent to NTU in the Ratio off mode.

General information

In no event will the manufacturer be liable for direct, indirect, special,
incidental or consequential damages resulting from any defect or
omission in this manual. The manufacturer reserves the right to make
changes in this manual and the products it describes at any time, without
notice or obligation. Revised editions are found on the manufacturer's
website.
Additional information
Additional information is available on the manufacturer's website.
Safety information
N O T I C E
The manufacturer is not responsible for any damages due to misapplication or
misuse of this product including, without limitation, direct, incidental and
consequential damages, and disclaims such damages to the full extent permitted
under applicable law. The user is solely responsible to identify critical application
risks and install appropriate mechanisms to protect processes during a possible
equipment malfunction.
Please read this entire manual before unpacking, setting up or operating
this equipment. Pay attention to all danger and caution statements.
Failure to do so could result in serious injury to the operator or damage
to the equipment.

Advertisement

Table of Contents
loading

Table of Contents