Hach 2100AN IS Basic User Manual page 12

Hide thumbs Also See for 2100AN IS:
Table of Contents

Advertisement

Available languages

Available languages

• In Calibration mode, automatic range and signal averaging on are
selected. When calibration is completed, all operational modes go
back to the last settings.
• All nephelometric (turbidity units of measure) calibrations are done at
the same time.
• Ratio-on and Ratio-off calibration data is measured and recorded at
the same time.
• The 4000-NTU and 7500-NTU standards do not have to be measured
during calibration if FNUs will be measured. Push CAL/Zero after the
StablCal calibration procedure
1. Push CAL/Zero.
2. Get the < 0.1 NTU
vial. Clean the vial with
The CAL/Zero light
a soft, lint-free cloth to
turns on, and the mode
remove water spots and
display shows "00". The
fingerprints. Do not
NTU value of the
invert the vial.
dilution water that was
used in the previous
calibration is shown on
the display.
12 English
1000 NTU standard is measured to complete the calibration
procedure.
• The 7500-NTU standard does not have to be measured during
calibration if turbidity less than 4000 NTU will be measured. Push
CAL/Zero after the 4000 NTU standard is measured to complete the
calibration procedure.
• The FNU values of StablCal standards and formazin standards are
calculated using the conversion factors of 1 FNU = 1 NTU.
3. Apply a small bead
4. Use the oiling cloth
of silicone oil from the
to apply the oil equally
top to the bottom of the
to the surface of the
vial.
vial. Remove the
excess oil. Make sure
that the vial is almost
dry.
5. Put the vial in the
6. Push ENTER.
sample cell holder with
The instrument display
the triangle on the vial
counts down, then
aligned with the
measures the standard.
reference mark on the
The next expected
sample cell holder.
standard (e.g., 20.00) is
Close the cover.
shown. The mode
display shows "01".

Advertisement

Table of Contents
loading

Table of Contents